| Model Number |
LC358XXL |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Provider alleges consumer has flipped back twice while in the chair.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been made available for evaluation.Should further information of the device become available, a follow-up report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
The alleged condition (flipping backwards) could not be duplicated during functional testing.Based on the visual inspection, the evaluator can not rule out improper back installation (post final release) per the owner's manual.(pgs.11-12).
|
| |
|
Event or Problem Description
|
|
Provider alleges consumer has flipped back twice while in the chair.
|
| |
|