BOSTON SCIENTIFIC CORPORATION WALLFLEX BILIARY; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00570890
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code Inflammation (1932)
Date of Event 12/01/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: approximated based on the date the manufacturer became aware of the event.The complainant was unable to provide the suspect device lot number.Therefore, the manufacture and expiration dates are unknown.(b)(4).The complainant indicated that the stent remains implanted and will not be returned for evaluation; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2021 that a wallflex biliary rx uncovered stent was implanted in the common bile duct to treat a malignant, extrinsic pancreatic mass during an endoscopic retrograde cholangiopancreatography (ercp) with stent placement procedure performed on an unknown date.Reportedly, the patient's anatomy was tortuous and was not dilated prior to the stent placement procedure.Five days post stent placement, the patient was admitted to the intensive care unit with cholangitis.Imaging was performed, and it was noted that the stent did not fully expand.Reintervention via ercp was performed and two plastic stents were implanted within the wallflex biliary stent to allow for biliary drainage.Reportedly, the physician was comfortable leaving the wallflex biliary uncovered stent implanted.The patient was reported to be fully recovered and the patient's current condition was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX BILIARY
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13176714
Report Number3005099803-2021-08148
Device Sequence Number4562607
Product Code FGE
UDI-Device Identifier08714729765332
UDI-Public08714729765332
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K140630
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00570890
Device Catalogue Number7089
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/14/2021
Initial Report FDA Received Date01/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-