| Model Number |
0220210000 |
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was loss of light.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of light.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: cib joey, rep, shutting off in middle of cases, sjc0015619 the failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be a bad main board and led module causing the light source to go to standby when activated.Additionally, poor air flow may affect ventilation of the light source unit.This may cause an intermittent loss of light.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|