| Model Number |
7K61-37 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer reported a falsely elevated architect b12 result on a patient.Results provided: (b)(6) 2021, sid (b)(6)= 364 pg/ml< repeated on (b)(6) 2021 = 274 pg/ml (both results using lot# 23108ui00).Normal range: 345-1485 pg/ml.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer reported a falsely elevated architect b12 result on a patient.Results provided: (b)(6) 2021, sid (b)(6)= 364 pg/ml< repeated on (b)(6) 2021 = 274 pg/ml (both results using lot# 23108ui00).Normal range: 345-1485 pg/ml.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
After further evaluation, the suspect medical device was changed to architect i2000sr, list 3m74 (irving, tx as manufacturing site) and submitted under manufacturer report number 3016438761-2022-00047-01.All further information will be documented under mdr number 3016438761-2022-00047-01.
|
| |
|