| Model Number |
A2025-080 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/13/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the 2.5x80mm armada 14 dilatation catheter was removed from the hoop coil without issue; however, the device could not be prepped as the balloon leaked saline.It was suspected that there was a pinhole in the balloon.Another balloon was used to complete the procedure.There was no patient involvement and no clinically significant delay.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for evaluation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
A visual and functional analysis was performed on the returned device.The reported balloon rupture was not confirmed; however, a leak within the shaft was observed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints.The investigation was unable to determine a cause for the leak within the shaft.It may be possible that the inner member/inflation lumen was inadvertently punctured with a guide wire or during handling/removal from the coil dispenser; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|