• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK LUER ACTIVATED VALVE; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK LUER ACTIVATED VALVE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C8612
Medical Device Problem Codes Failure to Disconnect (2541); Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device manufacturer address 1: (b)(6).The user facility submitted a medwatch report for this event: mw5105575.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the nurse pressed the retract button on a clearlink system extension set and it did not retract.The issue was identified during peripheral intravenous (piv) placement.The piv went sideways upon hitting the retract button and became stuck in the catheter.It took three hands to remove the needle from the catheter.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.A visual inspection was performed to the photograph using the naked eye which did not identify any abnormalities that could have contributed to the reported condition.The reported condition is not clearly observed via the provided photograph and due to the nature of the returned sample no additional testing could be performed.Therefore, the reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK LUER ACTIVATED VALVE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13177582
Report Number1416980-2021-07733
Device Sequence Number17299328
Product Code FPA
UDI-Device Identifier00085412070056
UDI-Public(01)00085412070056
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K112893
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2C8612
Device Lot NumberR21I18081
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/10/2021
Supplement Date Received by Manufacturer02/02/2022
Initial Report FDA Received Date01/06/2022
Supplement Report FDA Received Date02/03/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PIV/CATHETER
Patient SexUnknown
-
-