Model Number B2100-060
Medical Device Problem Codes Difficult to Remove (1528); Material Rupture (1546); Material Separation (1562)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 12/12/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event has been estimated.The customer reported the device is not returning.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.The additional device referenced in is being filed under a separate medwatch report number.
 
Event or Problem Description
It was reported that the procedure was to treat a heavily calcified re-thrombosed stent.After the 10x60mm armada 35 balloon dilatation catheter (bdc) was advanced to the lesion, and after several inflations of the balloon, between nominal pressure and above nominal, the balloon ruptured.It was noted the rupture occurred due to the previously implanted stent had fractured and caused the balloon to rupture on contact.During removal of the bdc resistance was met with the introducer and the central sheath of the balloon separated.It was noted another unspecified armada 35 bdc was advanced; however, the same exact issue occurred and bdc ruptured at an unknown pressure.Therefore, two fragments remained in the anatomy and the fragments were pressed against the vascular wall.There was no adverse patient sequela reported and no clinically significant delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record and complaint history of the reported lot could not be conducted because the lot number was not provided.The investigation determined the reported balloon rupture, difficulty removing the device, balloon separation, additional treatment and device embedded in the tissue or plaque appear to be related to circumstances of the procedure.It was reported by the account that the balloon interacted with the previously implanted fractured stent resulting in the reported balloon rupture.In addition, the resistance noted during removal and separation of the balloon material was likely the result of the ruptured balloon material catching on the introducer sheath during removal.Additional treatment was performed to press the separated fragments against the vascular wall.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13177583
Report Number2024168-2022-00170
Device Sequence Number4624677
Product Code LIT
UDI-Device Identifier08717648155529
UDI-Public08717648155529
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberB2100-060
Device Catalogue NumberB2100-060
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/13/2021
Supplement Date Received by Manufacturer01/13/2022
Initial Report FDA Received Date01/06/2022
Supplement Report FDA Received Date02/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-