• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS INC. PRELUDE IDEAL¿; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS INC. PRELUDE IDEAL¿; INTRODUCER, CATHETER Back to Search Results
Catalog Number PID6F16021PW/JPA
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/23/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event or Problem Description
The account alleges that following a successful pci procedure, the sheath was attempted to be removed.The clinician experienced difficulty removing the device and proceeded to create an incision in the skin and vessel for successful removal.No additional patient consequence to report.
 
Additional Manufacturer Narrative
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRELUDE IDEAL¿
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS INC.
1600 merit parkway
south jordan UT 84095
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS INC.
1600 merit parkway
south jordan UT 84095
Manufacturer Contact
bryson heaton
1600 merit parkway
south jordan, UT 84095
MDR Report Key13177585
Report Number1721504-2022-00001
Device Sequence Number17299329
Product Code DYB
UDI-Device Identifier00884450381892
UDI-Public884450381892
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Catalogue NumberPID6F16021PW/JPA
Device Lot NumberH2247525
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/24/2021
Supplement Date Received by Manufacturer01/25/2022
Initial Report FDA Received Date01/06/2022
Supplement Report FDA Received Date02/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-