Model Number FV770T
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 10/02/2017
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported miethke that a prosa sys w/adj.Dp.Unit a.Control res.(part # fv770t) was implanted during a procedure performed on (b)(6) 2016.According to the complainant, the valve was not able to be adjusted.The patient underwent a revision procedure on (b)(6) 2017.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: (b)(6).Weight: (b)(6).Height: 73 centimeters (cm).Gender: unknown.
 
Additional Manufacturer Narrative
This complaint was reopened in reference to (b)(4).Manufacturing site evaluation: manufacturing and quality control data: the progav® was manufactured by a qualified employee in (b)(6) 2016.Deviations during assembly did not occur.The valve was sterilized by miethke and released for shipment after final inspection.The progav® has a normal pressure range of 0 to 20 cmh2o.The valve was inspected as article fv770t.All parameters have been inspected and approved during the manufacturing process.All parameters have been assessed as meeting specification.Device examination: the valve was received submersed in an unidentified liquid in the provided returnkit.Visually, strong remnantsof tissue could be detected.Therefore, the condition of the product was rated with the highest risk level.Therefore, a device investigation was not possible due to the return condition.Result: a definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.However, manufacturing defects at the time of release have been excluded.The valve met all specifications of the final inspection when released from christoph miethke (b)(4).No further regulatory actions are required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROSA SYS W/ADJ.DP.UNIT A.CONTROL RES.
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key13177644
Report Number3004721439-2021-00838
Device Sequence Number4845529
Product Code JXG
UDI-Device Identifier04041906135553
UDI-Public4041906135553
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K120559
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2020
Device Model NumberFV770T
Device Catalogue NumberFV770T
Device Lot Number20027593
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2017
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/11/2017
Initial Report FDA Received Date01/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/26/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age1 YR
Patient SexUnknown
Patient Weight9 KG
-
-