| Model Number |
60728591 |
| Medical Device Problem Code |
Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/21/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the following: a white foreign body was found inside the syringe upon opening a new ssqk65/115vs syringe kit.No injury or adverse event was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Bayer product analysis received and examined the returned mr disposable syringe kit.Visual examination of the syringe barrel confirmed the presence of a white particle within the fluid path that was most likely material from the inner wall of the contrast spike luer which occurred during the component manufacturing/assembly process.A 12 month review of complaint history determined the complaint rate to be (b)(4) complaints per million (cpm).
|
| |
|