| Catalog Number |
06E50-68 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported (b)(6) prism htlv-i/htlv-ii results on multiple patients when performing a validation study on the alinity s analyzer.All of the (b)(6) prism samples tested negative on the alinity s.Western blot (wb) testing was (b)(6) for all but one sample, which was indeterminate (ind).Results provided: (b)(6) results (b)(6).No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.Patient identifier sids: (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
The evaluation of complaint data for the product and likely cause abbott prism htlv-i/htlv-ii identified normal complaint activity.As the lot numbers are unknown, the performance of the likely cause list number (6e50-68) was investigated by completing a review for non-conformances, potential non-conformances and deviations.This review did not identify any non-conformances, potential non-conformances or deviations for the list number.Manufacturing documentation for the likely cause did not identify any issues associated with the complaint issue.A review of overall customer field data was performed regarding initial reactive rate (irr), repeat reactive rate (rrr) and specificity for list number 6e50-68.The irr, the rrr and the specificity of the list number is within package insert specifications.A review of labeling concluded that the issue is sufficiently addressed.No systemic issue or product deficiency was identified.
|
| |
|
Search Alerts/Recalls
|