Catalog Number SGC0702 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/16/2021 |
Event Type
malfunction
|
Event Description
|
This is filed to report a leak.It was reported that during preparation of the steerable guide catheter (sgc), a loss of fluid column occurred.Therefore, the sgc was not used and was replaced.There was no clinically significant delay in the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.
|
|
Manufacturer Narrative
|
All available information was investigated.The reported leak (loss of fluid column) was not confirmed via returned device analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints reported from this lot.The investigation was unable to determine a cause for the reported leak.There is no indication of a product issue with respect to manufacture, design, or labeling.
|
|