• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CONTOUR VL; STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CONTOUR VL; STENT, URETERAL Back to Search Results
Model Number M0061801550
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/09/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a contour ureteral stent was used during a stent placement procedure , performed on (b)(6) 2021.During procedure, when the physician attempted to deploy the stent, it was noticed that the stent was ripped/torn.It was unknown how the procedure was completed.There were no patient complications reported as a result of this event.A photo of the complaint device was provided and showed the stent was torn.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR VL
Common Device Name
STENT, URETERAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13179884
Report Number3005099803-2022-00018
Device Sequence Number17576235
Product Code FAD
UDI-Device Identifier08714729058410
UDI-Public08714729058410
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K974541
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/22/2024
Device Model NumberM0061801550
Device Catalogue Number180-155
Device Lot Number0027717678
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/09/2021
Initial Report FDA Received Date01/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-