• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO MEDICAL CORPORATION DESTINATION RENAL GUIDING SHEATH; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO MEDICAL CORPORATION DESTINATION RENAL GUIDING SHEATH; INTRODUCER, CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Code Separation Problem (4043)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Date of Event 12/10/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Implanted date: device was not implanted.Explanted date: device was not explanted.Occupation: radiological technologist.The actual device has been returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.
 
Event or Problem Description
The user facility reported that during a right leg intervention via left common femoral approach, the destination sheath was stuck and when the sheath was pulled it separated.A cut down was performed, and the sheath was removed.The patient was closed; however, then returned for a left leg angiogram via the right common femoral artery because the below knee arteries had clotted off.The blood loss was greater than 250cc's.The patient was in stable condition.Additional information received 14-december-2021: the inner coil was exposed.The user facility stated they suspect that the access may have been through a previous self-expanding stent.
 
Additional Manufacturer Narrative
This report is being submitted as follow up no.1 to update section h3, and to provide the completed investigation results.One 7fr destination was returned for product evaluation.The returned sample included the sheath.An 8fr non-terumo sheath was also returned.The returned sample was subjected to visual analysis.The sheath was in two pieces.A piece not coil was also returned not inside any sheath.The first piece was found to be 39.7cm long.There were signs of stretching and flattening that began at 39.0cm from the sheath hub.The second piece was 7.5cm long.A slit was found on the shorter piece that run along most of its length.No coil was found present in the shorter sheath.The complaint can be confirmed for mechanical separation.The exact root cause cannot be determined.The likely root cause is that the destination sheath got caught on the stent.Review of dhr showed there were no issues noted during manufacture of the product that could have contributed to this complaint condition.Currently no action is recommended since this risk evaluation is within the predetermined limits in the fmea.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DESTINATION RENAL GUIDING SHEATH
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
Manufacturer (Section G)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
Manufacturer Contact
mary o'neill
950 elkton blvd.
elkton, MD 21921
8002837866
MDR Report Key13180102
Report Number1118880-2021-00221
Device Sequence Number5710461
Product Code DYB
UDI-Device Identifier00389701006192
UDI-Public00389701006192
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
K172995
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/01/2024
Device Model NumberN/A
Device Catalogue NumberRSR04
Device Lot Number0000051357
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/23/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2021
Supplement Date Received by Manufacturer01/25/2022
Initial Report FDA Received Date01/06/2022
Supplement Report FDA Received Date01/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
JET STREAM ATHERECTOMY BSCI
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age73 YR
Patient SexMale
Patient Weight84 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-