• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG EVIA DR-T; PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG EVIA DR-T; PACEMAKER Back to Search Results
Model Number 359529
Medical Device Problem Code Failure to Interrogate (1332)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/28/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this device was unable to be interrogated.Home monitoring showed the battery at 25 percent the same day.The device remains implanted.No adverse patient events were reported.Should additional information be received, this file will be updated.
 
Additional Manufacturer Narrative
The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Additional Manufacturer Narrative
This device was explanted and replaced on (b)(6) 2022.The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Additional Manufacturer Narrative
This device was explanted and replaced on (b)(6) 2022.The device was received and analyzed.Prior to the analysis, the manufacturing process for this device was re-investigated and all production steps were performed accordingly.There was no sign of any inconsistency during the manufacturing process.Particularly the final acceptance test proved the device functions to be as specified.Upon receipt, the device was not interrogatable, confirming the clinical observation.However, the memory content could be inspected.The inspection revealed an invalid memory content in an area related to specific device id data which resulted in the clinical observation.The pacemaker was subjected to an electrical analysis.The ability of the device to deliver therapies was verified.The anti-bradycardia pacing pulses proved to be normal and in amplitude and frequency as programmed.The pacemakers therapy capability was not compromised.During the further course of the analysis the pacemaker was reset.Subsequent interrogation was properly feasible.A stress test under different temperature environments was performed.The memory content remained valid.In conclusion, the clinical observation resulted from invalid memory content.The root cause for the invalid memory content was not determinable.However, external influences such as strong electromagnetic fields could be taken into consideration.After a manual correction of the invalid memory content the device was operating as specified.The analysis did not reveal any sign of a material or manufacturing problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIA DR-T
Common Device Name
PACEMAKER
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key13180208
Report Number1028232-2022-00081
Device Sequence Number5734791
Product Code NVZ
UDI-Device Identifier04035479101312
UDI-Public04035479101312
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P950037
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 12/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number359529
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/28/2021
Supplement Date Received by Manufacturer01/07/2022
01/31/2022
05/13/2022
Initial Report FDA Received Date01/06/2022
Supplement Report FDA Received Date01/11/2022
02/07/2022
05/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/23/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age79 YR
Patient SexMale
-
-