• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC SOLO HF CATHETER WITH SHERLOCK 3CG TIP POSITIONING SYSTEM; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC SOLO HF CATHETER WITH SHERLOCK 3CG TIP POSITIONING SYSTEM; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 12/20/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported "the power picc solo [left] behind a ¿piece¿ in the patient.I believe perhaps a piece of the stylet?.They did retrieve and save the ¿piece'.".
 
Additional Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of refv2191 showed no other similar product complaint(s) from this lot number.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and lot history review, applicable previous investigation(s), sample (if available), applicable manufacture records, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a broken stylet is confirmed; however, the exact cause is unknown.One 3cg stylet segment was returned for evaluation.An initial visual observation showed use residue on the returned sample.The stylet segment was observed to be of the distal tip of the stylet and was found to contain the epoxy tip.The stylet segment was found to be approximately 3.7 cm long.A slight bend was observed in the stylet approximately 0.3 cm distal to the break site.A microscopic observation revealed the break site in the polyimide tubing was uneven with a rough surface texture and plastic deformation.Some tapering, tearing, and slight delamination were observed in the polyimide tubing at the break site.The observed bend in the stylet segment was found to be between the first magnet interface distal to the break site.While the exact cause of the break in the returned stylet is unknown, possible causes include excessive manipulation of the stylet prior to or during placement and advancement/retraction of the stylet against resistance.
 
Event or Problem Description
It was reported "the power picc solo [left] behind a ¿piece¿ in the patient.I believe perhaps a piece of the stylet?.They did retrieve and save the ¿piece'.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC SOLO HF CATHETER WITH SHERLOCK 3CG TIP POSITIONING SYSTEM
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kelsey erickson
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13180584
Report Number3006260740-2021-05526
Device Sequence Number14880774
Product Code LJS
UDI-Device Identifier00801741034725
UDI-Public(01)00801741034725
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K091324
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number1395108QD
Device Lot NumberREFV2191
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/20/2021
Supplement Date Received by Manufacturer02/24/2022
Initial Report FDA Received Date01/06/2022
Supplement Report FDA Received Date02/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-