• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL MANUFACTURING IMPRESS®.BRAIDED CATHETER; PERIPHERAL CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL MANUFACTURING IMPRESS®.BRAIDED CATHETER; PERIPHERAL CATHETERS Back to Search Results
Catalog Number 56538MW2/JP
Medical Device Problem Codes Fracture (1260); Material Separation (1562)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 12/14/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event or Problem Description
The account alleges that during a percutaneous transluminal angioplasty [pta] stent graft procedure, the catheter tip detached within the patient.The physician had acquired right retrograde femoral artery access with a guidewire and a 5f catheter to successfully negotiated the patient's common iliac bifurcation in an attempt to deploy a stent graft within the patient's periphery.While attempting to withdraw the catheter from the patient, the catheter tip detached yet remained on the guidewire.The physician attempted to use a vascular snare device to externalized the tip when the tip detached again into two additional pieces.The foreign body remains within the patients right sfa.An additional procedure in under consideration.
 
Additional Manufacturer Narrative
The suspect device has been returned for evaluation.The product was examined visually.The complaint is confirmed.No definitive root cause could be determined.A search of the complaint database found no similar complaints for this lot number from the same customer.The device history record was reviewed, and no exception documents were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPRESS®.BRAIDED CATHETER
Common Device Name
PERIPHERAL CATHETERS
Manufacturer (Section D)
MERIT MEDICAL MANUFACTURING
14646 kirby dr
houston TX 77047
Manufacturer (Section G)
MERIT MEDICAL MANUFACTURING
14646 kirby dr
houston TX 77047
Manufacturer Contact
bryson heaton
1600 merit parkway
south jordan, UT 84095
8012084662
MDR Report Key13181471
Report Number3010665433-2022-00005
Device Sequence Number5935202
Product Code DQO
UDI-Device Identifier00884450009260
UDI-Public884450009260
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K053171
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/06/2023
Device Catalogue Number56538MW2/JP
Device Lot NumberE1755034
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/14/2021
Supplement Date Received by Manufacturer02/03/2022
Initial Report FDA Received Date01/06/2022
Supplement Report FDA Received Date02/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/03/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDEWIRE
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexUnknown
-
-