Model Number 6720-0330
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Pain (1994)
Date of Event 12/10/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
As reported: "pt presented with complaints of pain following multiple-dislocations.Pt had her hip replaced 'about a year ago'.Her hip has dislocated 'more than a half a dozen times since then'.Took culture-swabs to rule out infection.He suspects 'some sort of metalosis-reaction', following debridement of 'necrosed-looking tissue." a 46mm ha hemi shell, 35mm torx screw, 25 mm torx screw, "c" insert for a 28 mm head, and 28 + 8mm lfit head were revised.The stem is well-fixed and was left in-situ.Patient was revised to a 52e clusterhole shell with 3 screws, a 36e liner, and biolox 36 +7.5 mm head.Spoke to rep: surgeon originally suspected the patient suffered multiple dislocations due to shell malposition.However, intra-operatively, surgeon found signs of metal on metal wear and a possible allergic reaction or infection are thought to be possible.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIZE 3 ACCOLADE II 132 DEG
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13181947
Report Number0002249697-2022-00030
Device Sequence Number4257609
Product Code LPH
UDI-Device Identifier04546540664464
UDI-Public04546540664464
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K103479
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number6720-0330
Device Catalogue Number6720-0330
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/13/2021
Initial Report FDA Received Date01/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age63 YR
Patient SexFemale
-
-