Model Number MC2-5090S
Medical Device Problem Codes Fracture (1260); Mechanical Problem (1384)
Health Effect - Clinical Codes Failure of Implant (1924); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Event or Problem Description
Information was received that the rod did not fully achieve the desired distraction increments during lengthening sessions.X-rays revealed the right rod had a magnet fracture and a broken pin.The device was removed for final fusion.Additionally, it was reported that there was organic matter buildup around the rod and tissue discoloration.No patient adverse reaction or impact was reported.
 
Additional Manufacturer Narrative
During visual inspection, score marks consistent with expected markings due to incremental distraction were observed on the distraction rod.There was no visible cellulose buildup was found.The rod was functionally tested and unable to be distracted with manual distractor and electronic remote controller (erc).The rod was unable to be retracted with fast distractor to perform the stroke test.X-ray images of the internal components for the rod revealed broken internal parts such as; radial ball bearing, locking pin, retaining ring, anti-jam ring, and cylinder magnet.Sectioning of the housing tube revealed similar findings of the x-ray images.The radial ball bearing, locking pin, retaining ring, anti-jam ring and cylinder magnet were all broken.The reported failure mode was confirmed as the rod was not able to be distracted/retracted.There was no objective evidence to support the complaint description of tissue discoloration and cellulose buildup.After following up with the complaint reporter, no pictures or lab results were provided to support the complaint description of tissue discoloration and cellulose buildup.The iqc inspection data for the internal components; radial ball bearing, locking pin, retaining ring, anti-jam ring, and cylinder magnet were reviewed and the parts met design specifications per the engineering drawings.Review of the device history records revealed no discrepancies related to this complaint.The rod was manufactured in accordance with the specified requirements and met all of the required quality inspections.Based off the evidence seen from sectioning of the housing tube and the x-ray images of the rod, it can be deduced that the rod experienced repeated, excessive axial loading that caused serial failures of the various components that limit axial motion of the internal mechanism.
 
Event or Problem Description
No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS
101 enterprise #100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS
101 enterprise #100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise #100
aliso viejo, CA 92656
8583448112
MDR Report Key13182298
Report Number3006179046-2022-00107
Device Sequence Number2174958
Product Code PGN
UDI-Device Identifier00887517834041
UDI-Public887517834041
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K171791
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMC2-5090S
Device Lot Number8071101
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2021
Supplement Date Received by Manufacturer03/15/2022
Initial Report FDA Received Date01/06/2022
Supplement Report FDA Received Date03/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/11/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age12 YR
Patient SexFemale
Patient Weight32 KG
-
-