| Model Number |
MICL12.6 |
| Medical Device Problem Codes |
Off-Label Use (1494); Inadequacy of Device Shape and/or Size (1583); Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/05/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 12.6mm micl12.6 implantable collamer lens, -10.5 diopter into the patient's left (os) eye on (b)(6) 2021.The surgeon reports minimal vault.
|
| |
|
Additional Manufacturer Narrative
|
|
Weight, race, ethnicity: unk.Date of event: unk.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
B5-updated information: on (b)(6) 2022 the lens was exchanged with a longer lens.See mfr file# (b)(4), for replacement lens.The cause of the event is reported as a patient-related factor: "very unusual vault result.Wtw on pentacam was 11.3, topolyzer was 11.57, iolm700 was 11.6.Placed 12.6 but was considering 12.1.Vault low from the outset." h6-device code 1494: acd<3.0mm.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
D6b: corrected.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3- device evaluation: the lens was returned in liquid, wrapped in a cloth.Visual inspection found the haptic torn and debris throughout the lens.Dimensional inspection found the lens to be within specifications.Claim# (b)(4).
|
| |
|