| Model Number |
TMICL13.2 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Codes |
Intraocular Pressure Increased (1937); Pain (1994); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age, sex, weight, ethnicity, race: unk.Date of event: unk.(b)(4).Lens work order search: one similar complaint reported for units within the same lot.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 13.2mm tmicl13.2 implantable collamer lens -14.0/+3.5/162 (sphere/cylinder/axis) into the patient's left (os) eye on (b)(6) 2021.On (b)(6) 2021, the lens was explanted due to excessive vault, elevated iop, shallow ac and angles, pain, and vomiting.
|
| |
|
Additional Manufacturer Narrative
|
|
H3-device evaluation: the lens was returned in liquid in a vial.Visual inspection found that the haptic was torn.Claim# (b)(4).
|
| |
|
Search Alerts/Recalls
|