| Model Number |
3788 |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
12/20/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient lost therapy due to ipg being inoperable for years.Patient stopped charging the ipg for an extended period of time rending the ipg inoperable.As such, surgical intervention took place on (b)(6) 2021 wherein the ipg was explanted and replaced with a new ipg addressing the issue.Reportedly, therapy was confirmed post op.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is estimated.Exact event date is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
The patient reported failing to charge their implant properly to abbott.The device has not been returned for analysis.A review of documentation supplied with the implant states that the implant must be charged every 30 to 90 days to avoid battery depletion.Based on the information received, the cause of the reported incident is consistent with user error.
|
| |
|