MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 3501254BC
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 11/11/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a female patient underwent a breast augmentation surgery with implantation of a 125cc mentor memorygel breast implant prosthesis.Post-operatively, the patient was diagnosed with a rupture to the right breast prosthesis by an unknown methodology.As a result, the patient underwent removal at an undisclosed time.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: rupture.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On (b)(6) 2022, mentor completed an evaluation of the device utilizing the photo.Evaluation summary: according to the information received, the patient experienced a rupture in the breast implant.Upon visual evaluation of the image provided in the complaint, the implant was observed ruptured and inside of a plastic bag.As the product involved in this complaint was not received, the device couldn´t be analyzed according to our procedures.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On january 17, 2022, the mentor analysis lab received the medical device for evaluation.Furthermore, a patient identifier was provided and decontamination paperwork received with the device provided mentor with a date of explantation of (b)(6) 2021.An evaluation was complete on february 02, 2022.Visual analysis of the returned sample determined that the breast implant was found to have a tear on the anterior view, measuring 1.2 cm.Microscopic examination was performed on the edges of the rupture, and parallel striations were found in the whole area of the tear.Parallel striations are consistent with markings made by a sharp object perforating the implant shell.Based on the information currently available, microscopic examination of the returned product indicates that the implant could have been damaged during or subsequent to implantation.The product insert data sheet cautions to not allow cautery devices or sharp instruments, such as scalpels, suture needles, hypodermic needles, hemostats, adson forceps, or scissors to contact the device during the implantation or other surgical procedures.Reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
gabriel alfageme
3041 skyway circle north
irving, TX 75038
9497898687
MDR Report Key13184032
Report Number1645337-2022-00234
Device Sequence Number11810440
Product Code FTR
UDI-Device Identifier00081317000396
UDI-Public00081317000396
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/10/2025
Device Model Number3501254BC
Device Catalogue Number3501254BC
Device Lot Number9518340
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/13/2021
Supplement Date Received by Manufacturer01/10/2022
01/17/2022
Initial Report FDA Received Date01/06/2022
Supplement Report FDA Received Date01/11/2022
02/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/13/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-