MAUDE Adverse Event Report: MPRI SPECIFY SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI SPECIFY SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR Back to Search Results
Model Number 977C165
Medical Device Problem Code Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant products: product id: 97725, serial#: (b)(4), product type: external neurostimulator.Product id: 97715, serial#: (b)(4), implanted: (b)(6) 2022, product type: implantable neurostimulator.Product id: 977c165, serial#: (b)(4), implanted: (b)(6) 2022, product type: lead.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a trial patient who was using an external neurostimulator (ens) for unknown indications for use.It was reported that contacts 2 and 4 showed a bad connection no matter which channel it was plugged into in the wens and the ins.It was unknown if there were any contributing factors.The surgeon wiped the leads off with a damp ray tech and a dry one, changed the lead tails between the two channels in the wens and the ins and the bad connection followed the lead tail on the first lead.That lead was then swapped out and a new lead was used.This lead had the same bad connections on the same lead tail regardless of the channel in both the wens and ins.The surgeon continued to wipe and readjust and eventually contact 4 improved, but contact 2 remained with a bad connection.In the end this lead showed contacts 2 and 4 were normal, but contact 6 showed red (40 ,000).
 
Additional Manufacturer Narrative
Concomitant product id 97725 lot# serial# (b)(6) product type external neurostimulator product id 97715 lot# serial# (b)(6) implanted: (b)(6) 2022 product type implantable neurostimulator product id 977c165 lot# serial#(b)(6) implanted: (b)(6) 2022 product type lead medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information was received from the manufacturer representative (rep) reporting that the remaining high impedances on contact 6 did not resolve in post-op.The rep however did indicate that they were able to program around the high impedances to obtain effective therapy for the patient.The rep indicated that the lead that was swapped out was not returned yet but would be placed in the mail on 1/12/2022.The mail tracking number was provided.It was indicated that the provided information had been confirmed with the physician.No further complications were reported/anticipated.
 
Additional Manufacturer Narrative
H3: product analysis pli# 10 product id# 977c165 the returned device was subjected to a series of standard tests that include but is not limited to visual inspection and electrical testing.The returned device passed all testing in the laboratory and no anomalies were identified.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECIFY SURESCAN
Common Device Name
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13185481
Report Number2649622-2022-00367
Device Sequence Number16688114
Product Code LGW
UDI-Device Identifier00643169464599
UDI-Public00643169464599
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/26/2025
Device Model Number977C165
Device Catalogue Number977C165
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2022
Supplement Date Received by Manufacturer01/11/2022
03/09/2022
Initial Report FDA Received Date01/06/2022
Supplement Report FDA Received Date01/19/2022
03/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H10...."
Patient Age65 YR
Patient SexMale
-
-