| Catalog Number |
PDPB740D |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/14/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent a spinal fusion procedure on (b)(6) 2021 and suture was used.During the procedure, although no extra force was applied, three sutures detached from the needles in a row during interrupted suturing on the muscle layer.They were control release needles.There were no patient consequences reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint number: (b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This is a combination product, and the event has been reviewed for both the suture and the triclosan.Ingredient(s)- triclosan, dosage form - suture/solid/parenteral, strength - = 2360 ¿g /m.Related to: 2210968-2022-00160, and 2210968-2022-00161.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint number: (b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This is a combination product, and the event has been reviewed for both the suture and the triclosan.Ingredient(s)- triclosan, dosage form - suture/solid/parenteral, strength - = 2360 g /m.Related to: 2210968-2022-00160, and 2210968-2022-00161.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a spinal fusion procedure on (b)(6) 2021 and suture was used.During the procedure, although no extra force was applied, three sutures detached from the needles in a row during interrupted suturing on the muscle layer.They were control release needles.There were no patient consequences reported.Additional information was requested.
|
| |
|