Catalog Number RONYX22515UX |
Device Problems
Inadequacy of Device Shape and/or Size (1583); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Rupture (2208); Obstruction/Occlusion (2422); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/24/2021 |
Event Type
Injury
|
Event Description
|
It was reported that a resolute onyx coronary drug eluting stent was implanted and the patient heard the physician mentioning that stent was too small and that the left main artery was ruptured, than collapsed and a bypass may be needed.The patient has been in pain and her left arm and hand are frequently numb since the stent was implanted.Patient also reported heart pain, arteries spasm and continuing chest pain.The patient has attended the emergency room many times and has a follow up scheduled with a different physician.No further patient injury was reported.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: it was reported that a resolute onyx coronary drug eluting stent was intended to be used to treat the left main/ proximal lad.The device was inspected before use with no issues noted.The physician then reported that the correct sized stent was used, there was no artery rupture during the procedure and the lesion was not occluded and did not become occluded.The patient did not undergo bypass surgery.The physician reported that there are no adverse events related to the stent, and the reported adverse events and symptoms are not related to the procedure.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|