DEPUY SYNTHES PRODUCTS LLC KINCISE¿ RECHARGEABLE LI-ION BATTERY; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Model Number 1002-00-102
Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/20/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unspecified surgical procedure, while using the impactor device it was discovered that the battery device was not functioning properly.It was shorting out while pulling the trigger.It was further reported that while extracting the femoral head, the t-handle device cracked on the white handle where the surgeon grabs onto.According to the reporter, the surgeon was using a mallet on the t-handle and the white part cracked.It was reported that nothing broke off from the instrument.It was reported that there were no delays in the surgical procedure.It was not reported if a spare device was available for use.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: concomitant med products and therapy dates: impactor device, excel t-handle device, (b)(6) 2021.The surgeon¿s phone number was not provided.However, the reporter¿s name, phone number and email address were provided as (b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi: (b)(4).
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: h6: device history review: the actual device was not returned for evaluation; therefore, the reported condition was not confirmed.However, a device history review was performed which indicated that there were no non-conformances or abnormalities identified during the manufacture of the device that may have contributed to the reported condition.It was determined that the device passed all manufacturing and test requirements at the time of manufacture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KINCISE¿ RECHARGEABLE LI-ION BATTERY
Common Device Name
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
TECHTRONIC INDUSTRIES MEDICAL
1428 pearman dairy rd
anderson SC 29625
Manufacturer Contact
kara ditty-bovard
4500 riverside drive
palm beach gardens, FL 33410
6103142063
MDR Report Key13190832
Report Number1045834-2022-00025
Device Sequence Number19930856
Product Code GEY
UDI-Device Identifier00845384018537
UDI-Public00845384018537
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1002-00-102
Device Catalogue Number1002-00-102
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/20/2021
Supplement Date Received by Manufacturer02/14/2022
Initial Report FDA Received Date01/07/2022
Supplement Report FDA Received Date02/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/11/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-