| Model Number |
138107 |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation of the customer's complaint of an insufficient heat seal leading to a breach in sterility has confirmed the reported issue.Conmed received one 138107 returned unopened in original packaging.The lot numbers of the devices were verified via returned packaging.A visual inspection was performed the packaging was not sealed at one end.A functional inspection and dye leak testing were not necessary as the breach in sterility was an obvious failure.The manufacturing documents from the device history record have been reviewed and found no abnormalities that would contribute to this issue.A two-year lot history review shows this is the only complaint for this lot number and failure mode.A two-year review of complaint history revealed there has been a total of 99 complaints, regarding 7610 devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be 0.004.Conmed encourages the inspection and/or test of all medical equipment prior to use to ensure all devices are functioning as expected.The instructions for use (ifu) provides the user with information regarding proper care and use of this device.Per the ifu, the user is also advised to inspect and test each device before each use.This issue will continue to be monitored through the complaint system to assure patient safety.
|
| |
|
Event or Problem Description
|
|
During incoming inspection, the distributor rejected this device, catalog # 138107, lot 202006221, for a possible breach of sterility due to an insufficient heatseal.There was no contact with any patient as this was found during incoming inspection in (b)(6).The completion of the device evaluation confirmed an insufficient heatseal; therefore, this complaint meets the criteria for a reportable event.Due to the breach in sterility, this report is being raised as a malfunction with potential for injury upon reoccurrence.
|
| |
|
Search Alerts/Recalls
|
|
|