MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US S-ROM*SLEEVE PRX ZTT, 18D-LRG; S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US S-ROM*SLEEVE PRX ZTT, 18D-LRG; S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE Back to Search Results
Model Number 55-0524
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Necrosis (1971); Loss of Range of Motion (2032); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/07/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Left tha completed due to osteoarthritis.A depuy asr head, asr cup, and taper sleeve adapter were used.Patient completed of audible noise and popping in the hip.Revision completed secondary to adverse metal-on-metal reactions and cobalt chromium toxicity.Patient had moderate alval limited to the capsule with small amounts of tissue necrosis around the capsule.Doi: (b)(6) 2007; dor: (b)(6) 2020; left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Mfr# 1818910-2022-00391 is being retracted since it was found to be a duplicate of mfr# 1818910-2021-11399.Mfr# 1818910-2021-11399 will be kept for investigation purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13190910
Report Number1818910-2022-00391
Device Sequence Number4315081
Product Code LPH
UDI-Device Identifier10603295175292
UDI-Public10603295175292
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number55-0524
Device Catalogue Number550524
Device Lot Number2480099
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/20/2021
Supplement Date Received by Manufacturer02/01/2022
Initial Report FDA Received Date01/07/2022
Supplement Report FDA Received Date02/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/10/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ADAPTER SLEEVE 11/13 +0; ASR ACETABULAR CUPS 54; ASR UNI FEMORAL IMPL SIZE 47; SROM STM STD 36+12L 13X18
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexMale
-
-