|
The phenom 27 catheter was returned for analysis.No damage was found with the phenom 27 catheter hub.The catheter body was found kinked at 24.6cm and 22.7cm from the distal tip.No damage was found with the phenom 27 distal marker/tip.The phenom 27 catheter was flushed, water was seen exited under the strain relief and also from the distal tip.The strain relief was removed, the catheter body was found damaged (broken) at the distal end of the hub.An in-house 0.026¿ mandrel was inserted into the distal tip.Resistance was encountered at the kinked location but could not pass through the hub as the mandrel became stuck at the broken location.The phenom 27 catheter total length was measured to be 166.6cm, and the useable length was measured to be 160.2cm which is within specification (specification: 160cm ± 5).Based on the device analysis and reported information, the customer¿s report was confirmed.It is likely the damage (kinking) found with the returned phenom 27 catheter contributed to the event.The solitaire device used in the event was not returned; therefore, an analysis could not be performed and conformance to specification could not be assessed.Regarding the catheter break, the cause for the damage could not be determined.If information is provided in the future, a supplemental report will be issued.
|
|
Medtronic received a report that they could not pass the solitaire through the catheter as there was catheter occlusion.The patient was undergoing surgery for stroke mt treatment.The patient's blood flow was not indicated and vessel tortuosity was normal.It was reported that they could not pass the solitaire through the catheter as there was catheter occlusion. the catheter was flushed as indicated in the ifu.There were no patient symptoms or complications associated with the event.The device and accessory devices were prepared as indicated in the ifu.The product was replaced with a non-medtronic product. ancillary devices include a 8f terumo guidewire, a synchro 2, and a walrus bgc red 72 solitaire 6x40.
|