• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS 8000 CORE UNIT; CLINICAL CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS 8000 CORE UNIT; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Model Number COBAS 8000 CORE
Medical Device Problem Code Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/13/2021
Type of Reportable Event Malfunction
Event or Problem Description
There was a complaint of a questionable insulin result for 1 patient tested with a cobas 8000 core module.The questionable result was not reported outside the laboratory.The patient¿s repeated result was "cut in half" of the initial result.Specific values and units were requested, but not provided.The lot number and expiration date of the insulin reagent used were requested, but not provided.
 
Additional Manufacturer Narrative
The customer stated the qc was affected by the issue.The field service engineer (fse) determined that the pinch valve did not close properly.The fse replaced the pinch valve and ran calibration and qc checks.All checks met acceptance criteria.The investigation is ongoing.
 
Additional Manufacturer Narrative
An investigation of the alarm trace found 57 sample probe, 45 sample clot, and 106 sample short alarms.There were no relevant alarms at the time of the discrepant results.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS 8000 CORE UNIT
Common Device Name
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
HITACHI HIGH TECH CORP.
882 ichige hitachinaka
na
ibaraki 312-8 504
JA   312-8504
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key13191676
Report Number1823260-2022-00061
Device Sequence Number19934516
Product Code JJE
UDI-Device Identifier04015630928330
UDI-Public04015630928330
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K100853
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCOBAS 8000 CORE
Device Catalogue Number05641446001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2021
Supplement Date Received by Manufacturer01/14/2022
Initial Report FDA Received Date01/07/2022
Supplement Report FDA Received Date02/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-