• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US ARTHRO MOD RATCHET HDLE/CHUCK *EA; SURGICAL SCREWDRIVER, REUSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US ARTHRO MOD RATCHET HDLE/CHUCK *EA; SURGICAL SCREWDRIVER, REUSABLE Back to Search Results
Model Number 219215
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/20/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by sales rep that during an anterior cruciate ligament reconstruction surgery on (b)(6) 2021, it was observed that the two parts of the driver on the arthro mod ratchet hdle/chuck device separated leaving it useless and unable to be reattached.Nothing was left in the patient.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
Additional narrative: the lot number was unknown; therefore, the manufacturing site name was unknown.Udi: (b)(4).The lot number was unknown.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported by sales rep that during an anterior cruciate ligament reconstruction surgery on (b)(6) 2021, it was observed that the two parts of the driver on the arthro mod ratchet hdle/chuck device separated leaving it useless and unable to be reattached.Nothing was left in the patient.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
Additional narrative: the lot number was unknown; therefore, the manufacturing site name was unknown.Udi: (b)(4).The lot number was unknown.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-u medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the complaint device was received and evaluated.Visual observations reveals that the device has some marks of wear and use as usual on these type of reusable devices.The ratcheting chuck assembly is detached from its handle.At magnification of the shaft, the grip points were found damaged.As a potential cause cannot be associated to manufacturing, therefore a manufacturing record evaluation is not required.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection, this complaint can be confirmed.The root cause for this type of failure can be attributed to the repeated, rough and constant use of the device.Since this is a reusable device, the constant manipulation between surgeries and hard and tough tightening can lead to a chuck detachment.Per the ifu: end of useful instrument life is generally determined by wear or damage from handling or surgical use.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d9, h3, h6: the actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
SURGICAL SCREWDRIVER, REUSABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key13191763
Report Number1221934-2022-00061
Device Sequence Number19898050
Product Code LXH
UDI-Device Identifier10886705005505
UDI-Public10886705005505
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number219215
Device Catalogue Number219215
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/20/2021
Supplement Date Received by Manufacturer01/18/2022
01/29/2022
Initial Report FDA Received Date01/07/2022
Supplement Report FDA Received Date01/19/2022
01/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-