| Model Number |
219215 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/20/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by sales rep that during an anterior cruciate ligament reconstruction surgery on (b)(6) 2021, it was observed that the two parts of the driver on the arthro mod ratchet hdle/chuck device separated leaving it useless and unable to be reattached.Nothing was left in the patient.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional narrative: the lot number was unknown; therefore, the manufacturing site name was unknown.Udi: (b)(4).The lot number was unknown.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported by sales rep that during an anterior cruciate ligament reconstruction surgery on (b)(6) 2021, it was observed that the two parts of the driver on the arthro mod ratchet hdle/chuck device separated leaving it useless and unable to be reattached.Nothing was left in the patient.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional narrative: the lot number was unknown; therefore, the manufacturing site name was unknown.Udi: (b)(4).The lot number was unknown.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-u medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the complaint device was received and evaluated.Visual observations reveals that the device has some marks of wear and use as usual on these type of reusable devices.The ratcheting chuck assembly is detached from its handle.At magnification of the shaft, the grip points were found damaged.As a potential cause cannot be associated to manufacturing, therefore a manufacturing record evaluation is not required.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection, this complaint can be confirmed.The root cause for this type of failure can be attributed to the repeated, rough and constant use of the device.Since this is a reusable device, the constant manipulation between surgeries and hard and tough tightening can lead to a chuck detachment.Per the ifu: end of useful instrument life is generally determined by wear or damage from handling or surgical use.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d9, h3, h6: the actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.
|
| |
|