• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 WINGED INFUSION SET 20G X 0.75IN; NEEDLE, HYPODERMIC, SINGLE LUMEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 WINGED INFUSION SET 20G X 0.75IN; NEEDLE, HYPODERMIC, SINGLE LUMEN Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refu1294 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported patient contacted nursing team to advise them that her portacath needle line had fractured.Nursing team advised patient to clamp the line above where the fracture was located to ensure that nothing was able to enter the line.Patient booked to attend facility on the same day and the port needle was removed.Patient reviewed by the medical team.Patient's portacath was re-accessed safely, and their iv antibiotic treatment has resumed.
 
Event or Problem Description
It was reported patient contacted nursing team to advise them that her portacath needle line had fractured.Nursing team advised patient to clamp the line above where the fracture was located to ensure that nothing was able to enter the line.Patient booked to attend facility on the same day and the port needle was removed.Patient reviewed by the medical team.Patient's portacath was re-accessed safely, and their iv antibiotic treatment has resumed.
 
Additional Manufacturer Narrative
The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample analysis, applicable fmea documents, applicable manufacturing records, and labeling.Based on a review of this information, the following was concluded: the complaint of a breach in the infusion set was confirmed but the cause of the damage could not be determined from the photograph that was provided for investigation.The photo showed a winged infusion set that exhibited evidence of use.A stream of clear fluid was observed to exit from the extension set tubing at the distal end of the luer adaptor.The characteristics of the leak path in the tubing could not be discerned from the photo.While a breach in the tubing was confirmed, there were no distinguishing features that could aid in identifying a specific root cause.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WINGED INFUSION SET 20G X 0.75IN
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LUMEN
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13191821
Report Number3006260740-2021-05554
Device Sequence Number6045611
Product Code FMI
UDI-Device Identifier00801741065804
UDI-Public(01)00801741065804
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K863606
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number0604190
Device Lot NumberREFU1294
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/16/2021
Supplement Date Received by Manufacturer02/10/2022
Initial Report FDA Received Date01/07/2022
Supplement Report FDA Received Date02/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-