WRIGHTS LANE SYNTHES USA PRODUCTS LLC LOCKING SCREW FOR IM NAIL Ø 5MM/ 80MM/ XL25; SCREW, FIXATION, BONE Back to Search Results
Model Number 04.045.080
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/01/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional device product codes: hsb; jds.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2021, the patient was seen in clinic for 8 week follow up for repair of non union of distal femur fracture.Initial surgery date was (b)(6) 2021, where a distal femoral locking plate was removed and a rfna 5 degree nail was implanted.The most distal transverse lateral to medial screw in the rfna backed out and was tenting the skin.On (b)(6) 2021 removal of hardware was completed to take out the screw that backed out of the rfna nail.The remaining hardware was intact and left in place, with no further medical intervention planned at this time.This report is for one (1) locking screw for im nail ø 5mm/ 80mm/ xl25.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h6: photo investigation: the complaint device was not received for investigation.A photo investigation was performed based on the provided patient x-rays.Examination of the provided image was conducted.Based on the evidence observed in the images, it was found that the reported screw appears to be moved.The investigation has found sufficient evidence to confirm the reported event.A definitive assignable root cause could not be determined based on the provided information.As the device was not returned, an as-received condition could not be assessed.And a dimensional inspection document/specification review were not completed during the investigation no product design issues or discrepancies were observed (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW FOR IM NAIL Ø 5MM/ 80MM/ XL25
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
SYNTHES TUTTLINGEN
unter hasslen 5
tuttlingen 78532
GM   78532
Manufacturer Contact
kara ditty-bovard
1302 wrights lane east
west chester, PA 19380
6103142063
MDR Report Key13191889
Report Number2939274-2022-00053
Device Sequence Number20329498
Product Code HWC
UDI-Device Identifier10886982293039
UDI-Public(01)10886982293039
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K201336
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.045.080
Device Catalogue Number04.045.080
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/10/2021
Supplement Date Received by Manufacturer02/08/2022
Initial Report FDA Received Date01/07/2022
Supplement Report FDA Received Date02/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - NAILS: RFNA
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexFemale
-
-