• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC TORQUE LIMITING ATTACHMENT 1.5NM/QUICK COUPLING; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC TORQUE LIMITING ATTACHMENT 1.5NM/QUICK COUPLING; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Model Number 511.773
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on an unknown date, the torque limiter was found leaking brown fluid after sterilization.Patient involvement was unknown.This complaint involves one (1) device.This report is for (1) torque limiting attachment 1.5nm/quick coupling.This report is 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.The investigation could not be completed, no product was received; no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.A1.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Investigation summary: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot: part/lot combination is not valid, and device is not being returned, therefore, dhr could not be completed.If the device is returned or lot number can be confirmed, the dhr will be revisited.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.Service evaluation: product: 511.773 (torque limiter 1.5nm f/ao/asif) a visual and functional assessment was performed on the device.The following steps failed : section 70: check the torque.During service/repair, the following was observed (technician note): the minimum torque peak attained during evaluation was out of the specified limit for the part.During the service evaluation, the following failures were identified: functional: torque out of tolerance conclusion: reported condition is not confirmed , root cause: faulty parts , action : scrapping.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORQUE LIMITING ATTACHMENT 1.5NM/QUICK COUPLING
Common Device Name
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13192259
Report Number2939274-2022-00054
Device Sequence Number20133496
Product Code GEY
UDI-Device Identifier07611819128491
UDI-Public(01)07611819128491
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number511.773
Device Catalogue Number511.773
Device Lot Number18087
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/02/2022
Supplement Date Received by Manufacturer12/13/2022
12/13/2022
12/13/2022
Initial Report FDA Received Date01/07/2022
Supplement Report FDA Received Date02/08/2022
03/10/2022
12/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-