• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX BLUSELECT SUCTIONAID TRACHEOSTOMY TUBES; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX BLUSELECT SUCTIONAID TRACHEOSTOMY TUBES; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Model Number 101/800/080CZ
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the use of the product, leakage of air from the cuff was observed.No patient injury.No additional information is available.
 
Additional Manufacturer Narrative
One unit was returned for investigation.Upon physical inspection, it was found that the complained issue could be duplicated.Dhr review was not done because no fault was found.
 
Additional Manufacturer Narrative
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A product sample was received for evaluation.Visual and functional testing were performed.Under visual inspection the sample appeared to be in good condition.During manufacturing process the devices are 100% inflation tested, which includes inflating each device cuff and leaving it for a 12 hour period.Reductions in pressure over this time are considered a failure and the device would be rejected.The inflation test was repeated on the received sample.It was confirmed that after 12 hours the cuff was still fully inflated.Also no leak was found when inflated device was put under water.Based on results of testing the reported failure was not observed.No action was required since no fault was found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX BLUSELECT SUCTIONAID TRACHEOSTOMY TUBES
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
7-15-26 fukushima,
minneapolis, MN 55442
MDR Report Key13193586
Report Number3012307300-2022-00299
Device Sequence Number14963903
Product Code BTO
UDI-Device Identifier15019517076097
UDI-Public15019517076097
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K173384
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/25/2025
Device Model Number101/800/080CZ
Device Catalogue Number101/800/080CZ
Device Lot Number3933044
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/23/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/21/2021
Supplement Date Received by Manufacturer01/19/2022
04/25/2023
Initial Report FDA Received Date01/07/2022
Supplement Report FDA Received Date02/17/2022
05/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/10/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-