MAUDE Adverse Event Report: NEWPORT MEDICAL INSTRUMENT, INC. E360 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEWPORT MEDICAL INSTRUMENT, INC. E360 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number E360S-CN-NA
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/27/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic has not received the suspect device/component from the customer for evaluation nor has the device been evaluated by the service engineer.The engineer from a third-party maintenance company inspected the ventilator.The machine was restarted, and there was no abnormality after the repeated restarts.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that while in use on a patient this e360 ventilator generated an alarm and power disconnected message, the ventilator stopped delivering breaths to the patient and automatically shuts down.The patient was transferred to an alternate ventilator without injury.
 
Additional Manufacturer Narrative
Medtronic has not received the suspect device/component from the customer for evaluation nor has the device been evaluated by the service engineer.The engineer from a third-party maintenance company inspected the ventilator.The machine was restarted, and there was no abnormality after the repeated restarts.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that while in use on a patient this e360 ventilator generated an alarm and power disconnected message, the ventilator stopped delivering breaths to the patient and automatically shuts down.The patient was transferred to an alternate ventilator without injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
E360 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
NEWPORT MEDICAL INSTRUMENT, INC.
1620 sunflower ave
costa mesa CA 92626
Manufacturer (Section G)
NEWPORT MEDICAL INSTRUMENT, INC.
1620 sunflower ave
costa mesa CA 92626
Manufacturer Contact
kelly adams
2101 faraday avenue
carlsbad, CA 92008
7606035046
MDR Report Key13193699
Report Number2023050-2022-00001
Device Sequence Number4404602
Product Code CBK
UDI-Device Identifier10884521541078
UDI-Public10884521541078
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K101803
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberE360S-CN-NA
Device Catalogue NumberE360S-CN-NA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2021
Initial Report FDA Received Date01/07/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age72 YR
Patient SexMale
Patient Weight55 KG
Patient EthnicityNon Hispanic
-
-