| Model Number |
FX816T |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
Hydrocephalus (3272)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a m.Blue 5 sys w/ped.Control reservoir (part # (b)(4)) was implanted during a procedure performed on an unknown date.According to the complainant, the shunt system was believed to be operated in underdrainage.The patient underwent a revision procedure on an unknown date.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: (b)(6), height: unknown, weight: (b)(6), gender: female.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation: visual inspection: no significant deformations or damage of the valve were detected during the visual inspection.Permeability test: a permeability test has indicated that the valve has a blockage.Adjustment test: the m.Blue valve was tested and is not adjustable throughout the normal range.Klickforce and brake function test: the brake functionality test has shown that the brake function is operational, however the braking force cannot be measured due to the non-adjustability of the valve.Computer controlled test: not applicable.Internal inspection : after dismantling of the valve, deposits were found in the valve results: it should be noted that the product was received dry (i.E.Not submersed in liquid as recommended).The investigation of dry items is not significant due to the affect dry deposits of liquor and blood can have on product performance.In spite of this, we have investigated the valve to the best of our abilities.Based on our investigation results, we can determine clogging and non-adjustability of the valve at the time of our investigation.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|
| |
|
Search Alerts/Recalls
|