• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Model Number S-55-040-120-P6
Medical Device Problem Codes Difficult or Delayed Activation (2577); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/22/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat a greater than 70% stenosed vessel in the proximal superficial femoral artery (sfa).The vessel diameter was roughly 5-6mm in diameter and was prepared using a balloon at nominal pressures.Atherectomy was not used.The collagen plug was deployed for a non-abbott closure device and was causing blockage.The 5.5x40mm supera self expanding stent system (sess) was attempted to be deployed to treat the blockage; however, after a third to halfway of the stent was deployed the physician decided he didn't like how the non-abbott closure device was deforming the supera stent and decided it was not suitable after taking additional pictures.The physician removed the partially deployed supera stent under fluoroscopy with no issues and replaced with a covered stent that completed the procedure.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported difficulties were likely due to circumstances of the procedure.Based on the information provided, the non-abbott closure device was interacting with the supera stent causing the stent to deform.Reportedly, it was decided that the supera stent was not suitable after taking additional pictures.The physician removed the partially deployed supera stent under fluoroscopy with no issues and replaced with a covered stent that completed the procedure.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.D9/h3 device return status.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13196380
Report Number2024168-2022-00266
Device Sequence Number6045760
Product Code NIP
UDI-Device Identifier08717648226069
UDI-Public08717648226069
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P120020
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Model NumberS-55-040-120-P6
Device Catalogue NumberS-55-040-120-P6
Device Lot Number1050561
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/22/2021
Supplement Date Received by Manufacturer01/25/2022
Initial Report FDA Received Date01/07/2022
Supplement Report FDA Received Date01/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
7 FR SHEATH; ANGIO-SEAL
Patient Age64 YR
Patient SexFemale
Patient Weight74 KG
-
-