MICRO THERAPEUTICS, INC. DBA EV3 CONCERTO NYLON; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
|
Back to Search Results |
|
| Model Number |
NV-6-20-HELIX |
| Medical Device Problem Code |
Separation Failure (2547)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding 2 concerto coils that had resistance and a pusher was damaged and a third coil that wouldn't detach.The patient was undergoing a coil embolization procedure to treat an endoleak.Blood flow was normal.Vessel tortuosity was moderate.It was reported that all coils and accessory devices were prepared as indicated in the instructions for use (ifu).The 4x8 and 5x20 coils had resistance in the middle section of the microcatheter and accordioned.The 5x20 coil had damage to the distal end of the pusher but no damage was noted to the 4x8 coil.Continuous flush was not used but but the physician flushed the catheter before loading each coil, but the coils could not track through the catheter.The catheter was not damaged.The 6x20 coil could not be detached.Manual detachment was attempted twice without success; there was no attempt to detach the coil with an instant detacher.Another manufacturer's product was used to complete the procedure.There was no harm or injury to the patient.Ancillary device: terumo progreat 2.8 microcatheter.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received reported that the coils came out of the introducer sheath smoothly.It was noted that the catheter was not a medtronic product.
|
| |
|
Additional Manufacturer Narrative
|
|
As found condition: the concerto coil was returned for analysis within shipping box, within a biohazard pouch and within an opened concerto inner pouch.There was no instant detacher returned with the device.Visual inspection/damage location details: the pusher appeared to be separated on the proximal end; along with the ai, coupler tube, positive load indicator and break load indicator were missing and not returned.Therefore, any contributing factors could not be assessed.The concerto pushwire was found to be bent at 36.8cm to 37.7cm from proximal end.Under the microscope, the coin was found to be located against the lumen stop, the shield coil was found to be present, and the implant coil was still attached to the pusher.The implant coil was found damaged and stretched.No other anomalies were observed.Testing/analysis: the concerto pusher was not intact to concerto coil and due to its damaged condition, the concerto coil could not be used for testing with an in-house instant detacher.Conclusion: based on the device analysis and reported information, the customer¿s report of ¿non-detachment¿ was confirmed as the concerto implant coil was returned still attached to the pusher.In this event, the damages found with the pusher (broken/missing) likely contributed to the event causing the implant coil to not detach.Damage to the implant coil and pusher can occur if the user advances the coil against resistance.However, the cause for the damage could not be determined.There was no non-conformance to specification identified that led to the reported issues.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|