Model Number M006175263010
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/14/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a percuflex plus ureteral stent was used during a stent replacement procedure in the ureter, performed on "(b)(6) 2021".During the procedure and inside the patient, when the physician deployed the stent and removed the guidewire, it was noticed that a piece of the stent broke.No other device was used to grasp the broken piece of stent.Another percuflex plus was opened and successfully completed the procedure.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.A photo of the complaint device was provided and showed that a piece of the stent coil was detached/separated.
 
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Additional Manufacturer Narrative
Block h6: medical device problem code a0501 captures the reportable event of coil detachment of device or device component inside the patient.Block h10: the returned percuflex plus ureteral was analyzed, and a visual evaluation noted that the bladder pig tail was detached separate in 2 section.The suture string, the positioner and stabilizer did not return for the analysis.A photo was also attached in the complaint record.No other issues with the device were noted.The reported event was confirmed.According to the product analysis, it is possible that operational factors, such as interacting of the device with the suture string, manipulation or excess of force applied during it use, could have caused the detached observed.Consequently affect the performance of the device.Therefore, adverse event related to procedure is selected as the most probable root cause for the complaint.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Event or Problem Description
It was reported to boston scientific corporation that a percuflex plus ureteral stent was used during a stent replacement procedure in the ureter, performed on (b)(6) 2021.During the procedure and inside the patient, when the physician deployed the stent and removed the guidewire, it was noticed that a piece of the stent broke.No other device was used to grasp the broken piece of stent.Another percuflex plus was opened and successfully completed the procedure.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.A photo of the complaint device was provided and showed that a piece of the stent coil was detached/separated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCUFLEX PLUS
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13196991
Report Number3005099803-2022-00012
Device Sequence Number6071297
Product Code FAD
UDI-Device Identifier08714729302032
UDI-Public08714729302032
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K924608
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/02/2024
Device Model NumberM006175263010
Device Catalogue Number175-263-01
Device Lot Number0028248457
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/14/2021
Supplement Date Received by Manufacturer01/21/2022
Initial Report FDA Received Date01/07/2022
Supplement Report FDA Received Date02/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexFemale
Patient Weight54 KG
-
-