MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 14 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 14 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number A2020-120
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents and/or complaints reported from this lot.Based on the information provided, a conclusive cause for the reported leak could not be determined.It may be possible that the sidearm, inflation port or inflation lumen within the sidearm was damaged; however, without having the device to examine, a conclusive cause could not be determined.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a de novo lesion located in the peroneal artery that was heavily calcified, mildly tortuous and 60% stenosed.The armada 14 balloon catheter was advanced to the region of the distal middle third of the peroneal artery, and began to inflate, but pressure did not raise under fluoroscopy.It was observed that contrast was leaking at the hub.The device was removed and another armada was used to complete the procedure.There was no reported adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 14 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13197067
Report Number2024168-2022-00276
Device Sequence Number14964175
Product Code DQY
UDI-Device Identifier08717648155987
UDI-Public08717648155987
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K102705
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model NumberA2020-120
Device Catalogue NumberA2020-120
Device Lot Number1052641
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/20/2021
Initial Report FDA Received Date01/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDEWIRE COMMAND ES
Patient Age82 YR
Patient SexFemale
Patient Weight60 KG
-
-