• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON BD SHARPS COLL NEXT GEN 5.4QT RED 20/PACK; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON BD SHARPS COLL NEXT GEN 5.4QT RED 20/PACK; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number 305517
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/13/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturing location for this product is (b)(4).This site is an oem manufacturing site.Therefore, bd corporate headquarters in (b)(4) has been listed and the (b)(4) fda registration number has been used for the manufacture report number.Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event or Problem Description
It was reported bd sharps coll next gen 5.4qt red 20/pack had a broken lid.The following information was provided by the initial reporter, translated from (b)(6): "part of the lid is missing.".
 
Event or Problem Description
It was reported bd sharps coll next gen 5.4qt red 20/pack had a broken lid.The following information was provided by the initial reporter, translated from japanese: "part of the lid is missing.".
 
Additional Manufacturer Narrative
H.6.Investigation: pictures were received of the nonconforming product.According to the dhr review process, accordance with the dhr review process; it¿s not possible to get a result because the part number lot was not included in the reporting information for this customer complaint.A review of the ncmr¿s was performed; the result showed there were no issues reported like missing lids for the same part number throughout the last twelve months.According with this investigation, there isn¿t enough information like a lot number or pictures from the original packaging which is needed to rule out that this issue was generated due to a repackaging process.As part of this investigation, it was noted that this complaint was received from asia and as per customer information all the material sold to japan is being repackaging within a bd facility, therefore the evidence of the original packaging is needed to find the root cause of this issue.H3 other text : see h.10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SHARPS COLL NEXT GEN 5.4QT RED 20/PACK
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13197349
Report Number2243072-2021-03061
Device Sequence Number6046778
Product Code FMI
UDI-Device Identifier00382903055173
UDI-Public00382903055173
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K943141
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number305517
Device Catalogue Number305517
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2021
Supplement Date Received by Manufacturer04/21/2022
Initial Report FDA Received Date01/07/2022
Supplement Report FDA Received Date04/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-