DEVICOR MEDICAL PRODUCTS, INC MAMMOMARK; BIOPSY SITE IDENTIFIER Back to Search Results
Model Number MAM3008
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 12/14/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The mammomark biopsy site identifier is a sterile, single use device intended for use after an open surgical or percutaneous breast biopsy procedure to mark the biopsy site.The device is not available for analysis, which precludes a full investigation and analysis of the root cause.However, this failure mode is identified in the risk management file and will occur at the end of the procedure and involves the marker tip physically breaking.The failure mode involves the tip of a side deploy marker shearing on the cutter of a disposable probe and separating from the body of the marker.Should this failure mode occur, there is a chance the broken marker tip may be left in the patient's breast.Our current medical director reached out to the reporting physician to discuss the incident and stressed the importance of removal of the mammotome st device as delineated in ifu.The reporting physician was also provided a formal letter recommending the removal of the sheered materials.Based on the potential need of surgical intervention to remove the materials left behind we are submitting this medwatch report.
 
Event or Problem Description
During a procedure a marker was deployed into the patient but it did not show up in post images so they deployed a second marker.The second marker was successfully deployed but when the doctor tried to pull the product out from the breast there was resistance.After the procedure they could confirm that the distal tip of the applicator was still in the breast.The doctor is requesting information to see if the applicator tip needs to be removed.They are waiting on the results of the biopsy to see if the patient needs a lumpectomy and that they could remove the applicator tip at that time as well.
 
Additional Manufacturer Narrative
The mammomark biopsy site identifier is a sterile, single use device intended for use after an open surgical or percutaneous breast biopsy procedure to mark the biopsy site.The device is not available for analysis, which precludes a full investigation and analysis of the root cause.However, this failure mode is identified in the risk management file and will occur at the end of the procedure and involves the marker tip physically breaking.The failure mode involves the tip of a side deploy marker shearing on the cutter of a disposable probe and separating from the body of the marker.Should this failure mode occur, there is a chance the broken marker tip may be left in the patient's breast.Our current medical director reached out to the reporting physician to discuss the incident and stressed the importance of removal of the mammotome st device as delineated in ifu.The reporting physician was also provided a formal letter recommending the removal of the sheered materials.Based on the potential need of surgical intervention to remove the materials left behind we are submitting this medwatch report.
 
Event or Problem Description
During a procedure a marker was deployed into the patient but it did not show up in post images so they deployed a second marker.The second marker was successfully deployed but when the doctor tried to pull the product out from the breast there was resistance.After the procedure they could confirm that the distal tip of the applicator was still in the breast.The doctor is requesting information to see if the applicator tip needs to be removed.They are waiting on the results of the biopsy to see if the patient needs a lumpectomy and that they could remove the applicator tip at that time as well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAMMOMARK
Common Device Name
BIOPSY SITE IDENTIFIER
Manufacturer (Section D)
DEVICOR MEDICAL PRODUCTS, INC
300 e business way
unit 500
cincinnati OH 45241
Manufacturer (Section G)
DEVICOR MEDICAL PRODUCTS DE MEXICO
sor juanaines de la cruz
#20152 4-b
parque industrial,, tijuana 22440
MX   22440
Manufacturer Contact
vivian rodriguez
300 e business way
unit 500
cincinnati, OH 45241
MDR Report Key13197598
Report Number3008492462-2021-00011
Device Sequence Number5983948
Product Code NEU
UDI-Device Identifier00841911101659
UDI-Public(01)00841911101659(17)230126(10)F12129369D
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K082278
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/26/2023
Device Model NumberMAM3008
Device Catalogue NumberMAM3008
Device Lot NumberF12129369D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/14/2021
Supplement Date Received by Manufacturer12/14/2021
Initial Report FDA Received Date01/07/2022
Supplement Report FDA Received Date07/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-