• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 15750 ALTON PKWY RADIUS-7 KIT (BLUETOOTH ONLY); OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 15750 ALTON PKWY RADIUS-7 KIT (BLUETOOTH ONLY); OXIMETER Back to Search Results
Model Number 26945
Medical Device Problem Code Sensing Intermittently (1558)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional manufacuring narrative: attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
 
Event or Problem Description
The customer reported the device was intermittent.No consequences or impact to patient were reported.
 
Additional Manufacturer Narrative
Additional manufacuring narrative: other, other text: the returned device was evaluated.The device passed visual and functionality testing.Contamination was observed on the connector; however the complaint was not duplicated., corrected data: d11 concomintant added.
 
Event or Problem Description
The customer reported the device was intermittent.No consequences or impact to patient were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADIUS-7 KIT (BLUETOOTH ONLY)
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 15750 ALTON PKWY
15750 alton pkwy
irvine CA 92618
Manufacturer (Section G)
MASIMO - MEXICALI
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
matofa fautua
15750 alton pkwy
irvine, CA 92618
9498223661
MDR Report Key13197658
Report Number3019388613-2022-00013
Device Sequence Number6046796
Product Code DQA
UDI-Device Identifier00843997014267
UDI-Public00843997014267
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K193242
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 12/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number26945
Device Catalogue Number9916
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/16/2021
Supplement Date Received by Manufacturer01/25/2022
Initial Report FDA Received Date01/07/2022
Supplement Report FDA Received Date02/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/12/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BATTERY CHARGER
Patient SexUnknown
-
-