• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMRIS - DEERFIELD IMAGING, INC. IMRIS ORT100 OPERATING ROOM TABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMRIS - DEERFIELD IMAGING, INC. IMRIS ORT100 OPERATING ROOM TABLE Back to Search Results
Model Number ORT100
Medical Device Problem Codes Mechanical Jam (2983); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/09/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the rotational locking mechanism in an ort100 operating room table was jammed, such that it would not lock or unlock.The rotational lock is controlled through use of a foot pedal, which was difficult to depress and mechanically engage the rotational locking mechanism.No patient involvement was reported, the issue was observed during a qa check.
 
Additional Manufacturer Narrative
Imris engineer has visited the customer site and confirmed the rotational locking mechanism was seized.A followup report will be submitted once the investigation is complete.
 
Additional Manufacturer Narrative
Imris engineer has visited the customer site and confirmed the rotational locking mechanism was seized.A followup report will be submitted once the investigation is complete.
 
Event or Problem Description
It was reported the rotational locking mechanism in an ort100 operating room table was jammed, such that it would not lock or unlock.The rotational lock is controlled through use of a foot pedal, which was difficult to depress and mechanically engage the rotational locking mechanism.No patient involvement was reported, the issue was observed during a qa check.
 
Additional Manufacturer Narrative
The rotational locking mechanism is controlled by the user through a foot pedal at the base of the operating room table.An imris engineer found the foot pedal assembly was bent as a result of customer interaction with the mechanism, which is likely a result of the user attempting to lock the table position in place when the table rotational alignment was not correct at zero degrees.It is possible that fluid ingress from poor fluid management, or binding of foot pedal components, could have caused increased difficulty in using the foot pedal and led to deformation of the mechanism.The locking pin that mechnically locks the rotational movement was found to be intact.Preventive maintenance prior to this event did not observe difficulty in locking the table's rotational position or using the foot pedal.The foot pedal locking mechanism was replaced in the table and its function verified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMRIS ORT100 OPERATING ROOM TABLE
Common Device Name
OPERATING ROOM TABLE
Manufacturer (Section D)
IMRIS - DEERFIELD IMAGING, INC.
5101 shady oak road
minnetonka MN 55343 4100
Manufacturer Contact
paul campbell
5101 shady oak road
minnetonka, MN 55343-4100
7632036344
MDR Report Key13200899
Report Number3010326005-2022-00001
Device Sequence Number5125649
Product Code FQO
UDI-Device Identifier00857534006332
UDI-Public00857534006332
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberORT100
Device Catalogue Number114070-000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2021
Supplement Date Received by Manufacturer12/09/2021
Initial Report FDA Received Date01/08/2022
Supplement Report FDA Received Date06/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/25/2011
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-