• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RAZORCUT BLADE 4.5 DISP EP-1; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RAZORCUT BLADE 4.5 DISP EP-1; SAW, POWERED, AND ACCESSORIES Back to Search Results
Model Number 7205318
Medical Device Problem Code Flaked (1246)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/17/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that, during an arthroscopy surgery, the razorcut blade top was being shaved off, it created metal flake immediately after use.It is unknown if the pieces were retrieved from the patient.The procedure was successfully completed without delay using a back-up device.No further complications were reported.
 
Additional Manufacturer Narrative
H3, h6: the reported device was received for evaluation.A visual inspection revealed the device was not returned with original packaging and the device has debris on it.No physical damage is visible to the device.A functional evaluation of the returned device was powered up and ran with no evidence of metal flaking.Device passed all functional test.An analysis of the customer provided image shows top of a blade.It was determined the device did not contribute to the reported event.The complaint was not verified and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.A complaint history review concluded this was an isolated event.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.Per e-mail communication it is unknown if flakes from the blade were removed from the patient.The blade is an externally communicating device, it is neither manufactured nor intended for implantation.It is also not approved for long term internal tissue exposure and long-term implantation data is not available.The patient impact beyond the possible retained foreign body could not be definitively determined.If retained micro-motion or movement cannot be predicted.There is potential risk for tissue inflammation and/or pain.No further clinical/medical assessment is warranted at this time.Should additional clinical documentation become available in the future, the clinical/medical task may be re-evaluated.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAZORCUT BLADE 4.5 DISP EP-1
Common Device Name
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13202086
Report Number1219602-2022-00048
Device Sequence Number19930806
Product Code HAB
UDI-Device Identifier03596010251466
UDI-Public03596010251466
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/01/2025
Device Model Number7205318
Device Catalogue Number7205318
Device Lot Number50903201
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/18/2021
Supplement Date Received by Manufacturer02/11/2022
Initial Report FDA Received Date01/10/2022
Supplement Report FDA Received Date02/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-