Model Number 6260-9-336
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Pain (1994)
Date of Event 12/20/2021
Type of Reportable Event Serious Injury
Event or Problem Description
As reported: "right side hip revision, all components explanted were stryker components.The patient was re-implanted with all new revision components, and the case sheet is attached.The original surgery was done in 2011, the patient recently was reporting pain in his right hip, x-rays showed eccentric wear on the polyethylene component, as well as the femoral head on the trunnion.The patient¿s cobalt chrome levels were elevated, so the surgeon decided to revise the hip component.Adverse local tissue reactions were minimal, however the femoral head did have corrosion and fretting was noted on the trunnion.A portion of the superior lip of the polyethylene liner was fractured off.The locking mechanism was intact on the polyethylene/acetabular component junction, and the liner was not disassociated from the acetabular component.All components were removed." update: the shell was revised because the surgeon didn¿t like the position of the component.The damaged portion of the liner was the portion that had the lot code on it therefore it was illegible.
 
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V40 COCR LFIT HEAD 36MM/+10
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
joann lavatelli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13202843
Report Number0002249697-2022-00039
Device Sequence Number19930816
Product Code JDI
UDI-Device Identifier07613327032321
UDI-Public07613327032321
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K022077
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 01/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2016
Device Model Number6260-9-336
Device Catalogue Number6260-9-336
Device Lot NumberMKEPK9
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/20/2021
Initial Report FDA Received Date01/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/01/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2249697-05/07/2018-003R
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age75 YR
Patient SexMale
-
-