• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT PSL WITH PUREFIX HA 54MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT PSL WITH PUREFIX HA 54MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 542-11-54F
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Pain (1994)
Date of Event 12/20/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event or Problem Description
As reported: "right side hip revision, all components explanted were stryker components.The patient was re-implanted with all new revision components, and the case sheet is attached.The original surgery was done in 2011, the patient recently was reporting pain in his right hip, x-rays showed eccentric wear on the polyethylene component, as well as the femoral head on the trunnion.The patient¿s cobalt chrome levels were elevated, so the surgeon decided to revise the hip component.Adverse local tissue reactions were minimal, however the femoral head did have corrosion and fretting was noted on the trunnion.A portion of the superior lip of the polyethylene liner was fractured off.The locking mechanism was intact on the polyethylene/acetabular component junction, and the liner was not disassociated from the acetabular component.All components were removed." update: the shell was revised because the surgeon didn't like the position of the component.The damaged portion of the liner was the portion that had the lot code on it therefore it was illegible.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT PSL WITH PUREFIX HA 54MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
joann lavatelli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13202867
Report Number0002249697-2022-00043
Device Sequence Number19897860
Product Code LPH
UDI-Device Identifier07613327037203
UDI-Public07613327037203
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K143085
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2013
Device Model Number542-11-54F
Device Catalogue Number542-11-54F
Device Lot NumberJKMMTD
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/20/2021
Initial Report FDA Received Date01/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/08/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age75 YR
Patient SexMale
-
-