| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/14/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during an initial partial knee procedure, the provisional fractured during extraction.Attempts to obtain additional information have been made; however, no more is available at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was not confirmed.Visual inspection of the returned device found the device was not cracked or fractured.There was just some wear seen on the device.The device history record (dhr) was reviewed and no discrepancies were found.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon reassessment of the reported event, it was determined an mdr should not have been filed under the current mfr number.This event will be reported on 0001825034-2022-00502.
|
| |
|
Event or Problem Description
|
|
Upon reassessment of the reported event, it was determined an mdr should not have been filed under the current mfr number.This event will be reported on 0001825034-2022-00502.
|
| |
|